Test Code LAB6621 Varicella Zoster Virus PCR

Performing Section

Microbiology

Purpose/Principle

This assay is a real-time PCR system that allows the direct amplification and qualitative detection of VZV DNA from unprocessed specimens without nucleic acid extraction. The system consists of the Simplexa™ VZV Swab Direct assay, the LIAISON® MDX with software, the Direct Amplification Disc (DAD) and accessories. Fluorescent probes are used together with corresponding forward and reverse primers to amplify a well-conserved region of the VZV DNA polymerase gene and internal control targets, used to detect PCR failure or inhibition.

Special Collection Notes

Collect specimen according to standard operating procedure.

Specimen Type

Whole Blood

:

Purple Top- EDTA

 

CSF:

Sterile screw capped container

 

Standard Swab kit:

cutaneious or mucocutaneous lesions

Specimen Volume

-Whole Blood

Preferred: 3.0 mL

Minimum: 0.8 mL

 

-CSF

preferred:0.5 mL

Minimum : 0.3 mL

Processing Instructions

Store refrigerated and send sample on wet ice via courier.
If transport time is greater than 24 hours, freeze CSF or plasma at -20° Celsius.
Do not freeze whole blood.

Temperature

Ambient (unless >4 hour then Refrigerated)

Test Components

VZV PCR

Reference Range

Not Detected

TAT

48 Hours

Sample Stability

Ambient: 4 hours
Refrigerated: 72 hours

Additional Notes

Specimen will be rejected if on the incorrect swab or media.

Aliases

VZV PCR

CPT

87798

Test/Batch Schedule

Monday – Sunday

Method

Diasorin Simplexa