Test Code LAB2382 Manganese, Serum
Additional Codes
MNS
Reporting Name
Manganese, SUseful For
Monitoring manganese exposure using serum specimens
Nutritional monitoring
Performing Laboratory
Mayo Clinic Laboratories in RochesterSpecimen Type
SerumSpecimen Required
Patient Preparation: High concentrations of gadolinium, iodine, and barium are known to interfere with most metal tests. If gadolinium-, iodine, or barium-containing contrast media has been administered, a specimen should not be collected for 96 hours.
Supplies: Metal Free Specimen Vial (T173)
Collection Container/Tube: Plain, royal blue-top Vacutainer plastic trace element blood collection tube
Submission Container/Tube: 7-mL Mayo metal-free, screw-capped, polypropylene vial
Specimen Volume: 0.3 mL
Collection Instructions:
1. Allow the specimen to clot for 30 minutes, and then centrifuge to separate serum from the cellular fraction. Serum must be removed from cellular fraction within 4 hours of collection. Avoid hemolysis.
2. Remove the stopper. Carefully pour specimen into a Mayo metal-free, polypropylene vial, while avoiding transfer of the cellular components of blood. Do not insert a pipet into the serum to accomplish transfer, and do not ream the specimen with a wooden stick to assist with serum transfer.
3. See Metals Analysis Specimen Collection and Transport for complete instructions.
Specimen Minimum Volume
0.2 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Refrigerated (preferred) | 28 days | METAL FREE |
Ambient | 28 days | METAL FREE | |
Frozen | 28 days | METAL FREE |
Special Instructions
Reference Values
0-17 years: Not established
≥18 years: 0.5-1.2 ng/mL
Day(s) Performed
Tuesday
CPT Code Information
83785
Report Available
2 to 8 daysReject Due To
Gross hemolysis | Reject |
Gross lipemia | OK |
Gross icterus | OK |
Method Name
Triple-Quadrupole Inductively Coupled Plasma-Mass Spectrometry (ICP-MS/MS)