Sign in →

Test Code LAB10596 HIV-1 PCR, Quantitative

Aliases

HIV Quant
Viral Load

Special Collection Notes

Collect specimen according to standard operating procedure.

Specimen Type

Blood:
Purple Top Tube-EDTA (Spun Plasma)

Processing Instructions

Transport on wet ice. Specimen can be stored refrigerated (2C – 8C) for up to 24 hours prior to centrifugation.

Specimen Volume

Preferred: 6.0 mL
Minimum: 3.0 mL

Temperature

Refrigerated

Additional Notes

Specimen will be rejected if on the incorrect swab or media.

TAT

96 Hours

Test Components

HIV-1 PCR Result
HIV-1 PCR Value
HIV-1 PCR Value (Log10)

Reference Range

HIV-1 PCR Result

Not Detected

 

HIV-1 PCR Value

<=0.0 Copies/mL

 

HIV-1 PCR Value (Log10)

<=0.00 Log10 (copies/mL)

CPT

87536

Performing Section

Microbiology

Purpose/Principle

This test is intended only for viral load monitoring and should not be used for diagnostic confirmation of HIV-1 infection