Sign in →

Test Code LAB10144 Streptococcus Group A PCR

Aliases

STREPA
BSGA

Performing Section

Microbiology

Specimen Type

Tissue/Aspirate/Fluid/Other Sterile Specimen:

Sterile Container

 

Swab:

Aerobic Swab (E-Swab)

Specimen Volume

Aspirate/Fluid
Minimum:0.5 mL

Processing Instructions

Store and send ambient ASAP/STAT. Specimen must arrive to the lab within 24 hrs of collection.

TAT

72 hours

CPT

87651

Special Collection Notes

Collect specimen according to standard operating procedure.

Tissues, Aspirates, Fluids, and other Sterile Specimens: Collect specimen aseptically into a sterile specimen container.

Collect an aerobic culture swab using a COPAN ESwab. DO NOT collect with MRSA swab. Throat swabs are NOT acceptable.

Method

ABI 7500

Purpose/Principle

In this assay, DNA extracted from patient samples is subjected to real-time PCR amplification using
primers and a fluorescent probe directed against a target sequence in the Strep A genome (spy1258) (Kodani, et al 2011). The product is detected on an Applied Biosystems 7500 Taqman™ instrument using 5’-nuclease methodology. An exogenous positive control is co-amplified in each reaction to determine extraction efficiency and detect inhibition. Results are reported as Strep A Detected or Not Detected.

Temperature

Ambient

Sample Stability

Ambient: 24 hours

Test/Batch Schedule

Monday – Sunday

Test Components

Streptococcus Group A Qual

Reference Range

Streptococcus Group A Qual

Not Detected